Product info
- Product name:Anesthesia Machine
- Product code:CCA1014
Packaging Specifications
- Outer Packaging Size:0.7 × 0.67 × 1.3 m
- Gross Weight:82 kg
- Net Weight:0 kg
- Volume:0.61
Description
Main Body
- Material: High-strength engineering plastic rack, light, beautiful and corrosion resistant
- Scope of application: Adult, child
- Gas source:
- Flow meter:
- Flow rate of rapid oxygen supply: 25–75L/min
- Low oxygen pressure alarm: It will be alarmed when the oxygen pressure < 0.2MPa
- Vaporizer: It has the function of automatic compensation based on pressure, temperature and flow rate. The regulation range of evaporator concentration is 0–5vol%. Among Halothane, Enflurane, Isoflurane and Sevoflurane, one can be chosen for application as required by the customer.
- Respiratory circuit:
- Respiratory Bellows: Bellows for adults, bellows for children; tidal volume range: 0–1500ml
Ventilator
Display & Mode
- Display mode: High-definition 5.7" LCD screen display
- Ventilation mode: IPPV, SIPPV, SIMV, MANUAL
- Ventilation functions: PEEP, SIGH, IRV
Ventilation Parameters
- Tidal volume: 50–1500ml
- Inspiratory trigger pressure: -10–20cmH₂O
- Rate: 2–99bpm
- SIMV rate: 2–20bpm
- PEEP: 2–20cmH₂O
- I:E: 2:1–1:8
- Pressure limitation: 10–60cmH₂O
- SIGH: 1.5 times inspiratory time/100bpm
Monitoring Parameters
Parameters for ventilation monitoring: tidal volume, minute ventilation volume, IPPV rate, SIMV rate, total respiratory rate, I/E, peak pressure of airway pressure, time waveform, flow rate - time waveform, PEEP, inspiratory trigger pressure, average pressure, oxygen concentration, battery capacity
Security Alert System
Oxygen Concentration Alarm
- Upper limit setting range: (21%–99%)
- Lower limit setting range: (18%–80%)
Airway Pressure Alarm
- Upper limit setting range: 5–60cmH₂O
- Lower limit setting range: 0–50cmH₂O
Minute Ventilation Volume Alarm
- Upper limit setting range: 1–20L/min
- Lower limit setting range: 0–19L/min
Tidal Volume Alarm
- Upper limit setting range: 100–1500ml
- Lower limit setting range: 0–1400ml
Sustained High-pressure Alarm
It will give alarm when stress have consistently been higher than 25cmH₂O
Other Alarms
- Suffocation alarm
- Power alarm
- Intubation off alarm
- Battery voltage alarm
Power
- AC 220V 50Hz (UPS, Storage battery)
Optional Items
- Mechanical Arms
- End-tidal CO₂ monitoring
- Specific interface of waste anesthesia gas